Čas odozvy:
                                            
                                            
                                        
									    90 % in less than 90 seconds
                                    
                                
    						
								
                                    
                                        
                                            
                                                Detekčný limit:
                                            
                                            
                                        
									    0.005 mg/L HOCl
                                    
                                
    						
								
                                    
                                        
                                            
                                                Hmotnosť:
                                            
                                            
                                        
									    6,5 kg 14,3 lbs
                                    
                                
    						
								
                                    
                                        
                                            
                                                Interferencie:
                                            
                                            
                                        
									    No interference from chloramines.
Chlorine dioxide and ozone interfere with measurement.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Interval merania:
                                            
                                            
                                        
									    Continuous
                                    
                                
    						
								
                                    
                                        
                                            
                                                Interval údržby:
                                            
                                            
                                        
									    Measurement cell: 6 months for membrane and electrolyte for typical application
pH: 1 - 1.5 years for typical application
                                    
                                
    						
								
                                    
                                        
                                            
                                                Kalibračná metóda:
                                            
                                            
                                        
									    Zero Calibration: electrically (automatic) or with de-chlorinated water
Calibration: Comparison of lab method with process sample
pH Calibration: Single- or two-point using standards or comparison of lab method with process sample
                                    
                                
    						
								
                                    
                                        
                                            
                                                Kalibračný interval:
                                            
                                            
                                        
									    2 mesiacov
                                    
                                
    						
								
                                    
                                        
                                            
                                                Krytie:
                                            
                                            
                                        
									    IP66 / NEMA 4X
                                    
                                
    						
								
                                    
                                        
                                            
                                                Obsah balenia:
                                            
                                            
                                        
									    9184sc Chlorine Sensor Panel, Membranes and Electrolyte and Manual. Does not include pH sensor.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Odvod vzorky:
                                            
                                            
                                        
									    0.5" inner diameter
                                    
                                
    						
								
                                    
                                        
                                            
                                                Parameter:
                                            
                                            
                                        
									    Active Free Chlorine: Hypochlorous Acid Only
                                    
                                
    						
								
                                    
                                        
                                            
                                                Podmienky skladovania:
                                            
                                            
                                        
									    -20 °C - 60 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Popis:
                                            
                                            
                                        
									    9184sc Hypochlorous Acid Chlorine Amperometric Sensor
                                    
                                
    						
								
                                    
                                        
                                            
                                                Presnosť:
                                            
                                            
                                        
									    2 % or ±10 ppb HOCl, whichever is greater at pH <8
                                    
                                
                                    
                                        
                                            
                                            
                                                 
                                            
                                        
									    2 % or ±10 ppb TFC, whichever is greater at pH <7.5
                                    
                                
    						
								
                                    
                                        
                                            
                                                Prevádzková teplota:
                                            
                                            
                                        
									    0 - 45 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Prietok:
                                            
                                            
                                        
									    0.25 L/h (200 - 250 mL/min) auto--regulated by flow through cell
                                    
                                
    						
								
                                    
                                        
                                            
                                                Princíp merania:
                                            
                                            
                                        
									    Amperometric/Membrane (electrode, membrane, electrolyte)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Relatívna vlhkosť:
                                            
                                            
                                        
									    0 - 90 % non-condensing
                                    
                                
    						
								
                                    
                                        
                                            
                                                Rozsah merania:
                                            
                                            
                                        
									    0 - 20 mg/L hypochlorous acid HOCl
                                    
                                
    						
								
                                    
                                        
                                            
                                                Rozsah pH:
                                            
                                            
                                        
									    4 - 8, acidification unit available for pH > 8
                                    
                                
    						
								
                                    
                                        
                                            
                                                Rozsah tlaku:
                                            
                                            
                                        
									    0.1 - 2 bar in flow cell
                                    
                                
    						
								
                                    
                                        
                                            
                                                Spôsob inštalácie:
                                            
                                            
                                        
									    Flat, vertical surface (panel, stand, etc.)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Standard deviation:
                                            
                                            
                                        
									    0.7 %
                                    
                                
    						
								
                                    
                                        
                                            
                                                Teplota vzorky:
                                            
                                            
                                        
									    2 - 45°C (35,6 - 113°F)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Teplotní kompenzace:
                                            
                                            
                                        
									    Automatic over sample temperature range
                                    
                                
                                    
                                        
                                            
                                            
                                                 
                                            
                                        
									    Automatic over sample temperature range
                                    
                                
    						
								
                                    
                                        
                                            
                                                Vstup vzorky:
                                            
                                            
                                        
									    0.25" outer diameter
                                    
                                
    						
								
                                    
                                        
                                            
                                                Záruka:
                                            
                                            
                                        
									    24 mesiacov