EZ7676 Total Nitrogen + Total Phosphorus Analyser

EZ7676 Total Nitrogen + Total Phosphorus Analyser
Kat. číslo produkta: EZ7676.XXXXXXXX
Jednotková cena Kontaktujte nás

Obsah balenia

EZ7676 Total Nitrogen & Total Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (HCl, NaOH and H₂SO₄ Solution) and 3 x empty 2.5L Glass Reagent Containers with Fittings (Colour, Reductor & Persulphate Solution)

Technické údaje

Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analógové výstupy: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Automatické čistenie: Áno
Č. prúdov vzoriek: 1, 2, 4, or 8
Certifikácia: CE compliant / ETL certified
Demineralised water: Consumption: 140 mL/analysis (hour)
Detekčný limit: TN: ≤ 4 mg/L
TP: ≤ 0.5 mg/L
Digitálne výstupy: Optional: Modbus (TCP/IP, RS485)
Doba cyklu : Standard measurement cycle time for both TN & TP: 60 minutes
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Hmotnosť: 25 kg
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencie: TN: The main interferences are Br- and I-. When the amount of I- is 2.2 fold of the amount of TN, or the amount of Br- is 3.4 fold of the amount of TN, this will interfere on the test results. Dissolved organic matter, surfactants and Chromium (VI) interfere. Various inorganic substances not normally found in natural water, such as Chlorite [ClO₂-] and Chlorate [ClO₃-], may interfere.

TP: Arsenic (V), Chromium (VI), Copper (II) > 10 mg/L, Iron (III) > 10 mg/L, Sulphide > 2 mg/L, and Vanadium (V), Silica > 60 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Kalibrácia: Automatic, 2-point; frequency freely programmable
Kvalita vzorky: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Materiál: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metóda merania: TN: UV photometric measurement at 220 nm after persulphate digestion in alkaline medium, based on APHA 4500-NO3 (B)
TP: Colorimetric measurement at 700 nm using ascorbic acid reduction and molybdate colour solution after persulphate digestion in acidic medium, based on APHA 4500-P
Napájanie: 230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Obsah balenia: EZ7676 Total Nitrogen & Total Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (HCl, NaOH and H₂SO₄ Solution) and 3 x empty 2.5L Glass Reagent Containers with Fittings (Colour, Reductor & Persulphate Solution)
Odtok: Atmospheric pressure, vented, min. Ø 32 mm
Parameter: Total Nitrogen (TN), Total Phosphorus (TP)
Presnosť: Better than 3% (TN) and 2% (TP) full scale range for standard test solutions
Reagent Requirements: Keep between 10 - 30 °C
Rozmery (V x Š x H): 690 mm x 465 mm x 330 mm
Rozsah merania: 4 - 100 mg/L TN; 0.5 - 50 mg/L TP
Rýchlosť prietoku vzorky: 100 - 300 mL/min
Teplota prostredia: 10 - 30 °C, ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Teplota vzorky: 10 - 30 °C
Trieda ochrany: Analyser cabinet: IP55 / Panel PC: IP65
Validation: Automatic; frequency freely programmable
Vzorka: tlak: By external overflow vessel
Záruka: 24 mesiacov