Detekčný limit:
TOC: 0.6 mg/L C with Automatic Range Selection
TN: 0.4 mg/L N with Automatic Range Selection
TP: 0.4 mg/L P with Automatic Range Selection
Displej: LCD (podsvietený):
High contrast 40 character x 16 line backlit LCD with LED backlight
Doba cyklu :
From 10 minutes, depending on range and application
Eexp / Nebezpečné miesto:
Certification options are available to European Standards (ATEX Zone 2 - maximum T3 for TP analyser) and North American Standards (Class I Division 2)
Expected Measuring Range:
0 - 500 mg/L
Hmotnosť:
90 - 120 kg
Enclosure weight may change depending on system optional features.
Jazyk užívateľského rozhrania:
Slovak
Komunikace: digitální :
Modbus RTU, Modbus TCP/IP & Profibus
(when the Profibus option is selected, the digital output signals are sent through the Profibus converter with its specific communication protocol)
Krytie:
IP44; volitelný IP54 s profouknutím vzduchem
Metóda merania:
TOC: NDIR measurement of CO₂ after oxidation
TN: direct photometric analysis of Nitrate after oxidation
TP: colorimetric analysis of Phosphate with standard Vanadomolybdophosphoricacid method after oxidation
Multi-Stream:
Valves for up to 3 streams with up to six 4-20 mA signals
The number of available outputs depends on the manual stream configuration.
Objem vzorky:
Up to 8.0 mL
Obsah balenia:
B7000 TOC/TN/TP Analyser, Tubing, Fuses, Ferrules, Drain, Acid, Base & TN Dip Tubes, CO₂ Filter & B7000 TOC/TN/TP User Manual
Odtok:
Typically ambient (for applications with high drain pressure, optional systems are available)
Opakovateľnosť:
TOC: ±3% of reading or ±0.3 mg/L C, whichever is greater
TN: ±3% of reading or ±0.2 mg/L N, whichever is greater
TP: ±3% of reading or ±0.2 mg/L P, whichever is greater
Oxidační metoda:
Innovative Two-Stage Advanced Oxidation Process (TSAO) using Hydroxyl Radicals
Parameter:
Direct measurement of TOC, TIC, TC, TN, TP; COD, BOD via correlation; VOC via calculation
Počet kanálov:
1 Channel (One 4-20mA output is indluded as standard)
Požiadavky na napájanie (Hz):
50 Hz
Požiadavky na napájanie (V):
230 V AC
Přípustný rozsah chloridů :
Up to 30%
Rozsah merania:
0 - 500 mg/L C/N/P
Servisný interval:
6 months service intervals
Teplota prostredia:
5 - 40 °C
Teplota vzorky na vstupe:
2 - 60 °C
Ukladanie dát:
Previous 9999 analysis data on screen in the microcontroller memory and storage of data archive for the lifetime of the analyser in the SD/MMC card.
Previous 99 fault data on screen in the microcontroller memory and storage of fault data archive for the lifetime of the analyser in the SD/MMC card.
Užívateľské rozhranie:
Microcontroller with membrane keyboard
Velikost částic:
Up to 2 mm, soft particulates
Vlhkosť:
5 - 85 % (non-condensing)
Voľba rozsahu:
Automatic or Manual Range Selection
Výstup:
One programmable 4-20 mA analogue output signal (typically for TOC)
Maximum impedance: 500 ohms
For systems requiring more than six 4-20 mA standard outputs, 4-20 mA Output Multiplex option is implemented to provide 4-20 mA data for up to 35 output signals
Záruka:
12 mesiacov
Zdroj napájania:
TIC/TOC & TN/TP - 230V